18
immunity
test
iEc 60601
test level
compliance
level
Electromagnetic environment –
guidance
conducted rF
iEc 61000-4-6
3 V rms
150 kHz
to 80 MHz
Not
applicable
Portable and mobile RF communications
equipment should be used no closer to
any part of the TH30F, including cables,
than the recommended separation
distance calculated from the equation
applicable to the frequency of the
transmitter.
recommended separation distance
d = 1,2 √P
d = 1,2 √P 80 MHz to 800 MHz
d = 2,3 √P 800 MHz to 2,5 GHz
where P is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter manufacturer
and d is the recommended separation
distance in meters (m).
Field strengths from fixed RF transmitters,
as determined by an electromagnetic
site survey, a should be less than the
compliance level in each frequency range
b.
Interference may occur in the vicinity of
equipment marked with the following
symbol:
rF rayonnée
iEc 61000-4-3
3 V/m
80 MHz
à 2,5 GHz
3 V/m
notE 1:
At 80 MHz and 800 MHz, the higher frequency range applies.