5-4

Medical Vacuum Filter Modules

4107 9022 25.00

9.  Purchase filter service kit from BeaconMedæs

The  kit  will  come  with  the  filter  element 

(including  double  O-rings  preinstalled)  and 

filter  housing  O-ring.  Each  filter  housing  has 

a  label  specifying  MV80  or  MV160.  For  an 

MV80  filter,  the  kit  P/N  is  4107  4018  70.  For 

an MV160 filter, the kit P/N is 4107 4018 71. 

More information can be found by referencing 

the Parts List. Please contact BeaconMedæs to 

have  the  Parts  List  sent  to  you  or  download 

an  electronic  version  from  the  website  at 

 

www.beaconmedaes.com.

6.  The  picture  above  shows  the  filter  housing 

removed and the filter element exposed.

7.  Remove  the  filter  element  by  giving  a  gentle 

pull downwards and dispose of the filter using 

hospital procedure for biohazard waste. 

8.  Dispose of gloves. Put on a new pair of gloves 

before proceeding further.

Summary of Contents for 4107 4018 70

Page 1: ...Installation Operation and Maintenance Instructions Medical Vacuum Filter Modules Part number 4107 9022 25 00 24 July 2018...

Page 2: ......

Page 3: ...l number Serial number Option s included Any information service or spare parts requests should include the machine serial number and be directed to BeaconMed s 1059 Paragon Way Rock Hill SC 29730 Tel...

Page 4: ...Upon Receiving 2 1 2 2 Handling 2 1 2 3 Location 2 1 2 4 Space Requirements 2 1 2 5 Inlet and Outlet Piping 2 1 2 6 Floor Mounting Instructions 2 1 2 7 Drain Flask Installation Instructions 2 1 3 0 O...

Page 5: ...ccor dance with requirements stated by NFPA 99 ISO 7396 1 HTM 02 01 and all applicable local codes 1 5 Environmental Declarations Disposal General When developing products and services Bea conMed s tr...

Page 6: ...contaminated substances and materials separately according to local applicable environ mental legislations Dispose all components according to the appli cable disposal regulations 1 2 Medical Vacuum F...

Page 7: ...ndoors in a clean well ventilated environment Areas of excessive dust dirt or other air borne particulate should be avoided 2 4 Space Requirements The MV Filter Modules should be placed to ensure easy...

Page 8: ...ow decals pictured above may or not be included If so discard They are not required for this application 5 Lastly install the drain flask by screwing it onto the ball valve located on the bottom of th...

Page 9: ...cedure In order to comply with ISO code one filter must be in reserve at all times This means that the but terfly valves on both sides of one filter must be closed at all times 3 4 HTM 02 01 2006 Oper...

Page 10: ...lockages in system See maintenance schedule in section 5 2 See operational procedures in sections 3 2 3 4 for the minimum required filters to be offline Inspect tubes for blockages Reduced vacuum leve...

Page 11: ...ist Please contact BeaconMed s to have the Parts List sent to you or downloadanelectronicversionfromthewebsiteat www beaconmedaes com In addition to the PPE kit below are the recommended gloves to be...

Page 12: ...nded to replace the drain flask with every filter element change The drain flask is available through BeaconMed s The P N for the flask is 4107 6558 95 More information can be found by referencing the...

Page 13: ...lter assembly and ease in removing the filter housing 4 Unscrew the filter housing A strap wrench will ease in loosening the filter housing After the filter housing is loosened it can be unscrewed by...

Page 14: ...ilter the kit P N is 4107 4018 71 More information can be found by referencing the Parts List Please contact BeaconMed s to have the Parts List sent to you or download an electronic version from the w...

Page 15: ...eated 11 Install new filter housing O ring that is provided with kit and re install housing by fully tightening the housing onto the head 12 Re install the drain flask by screwing it onto the ball val...

Page 16: ...6558 95 Replace with every filter element change Differential Pressure Gauge 1629 0438 00 Check monthly If the gauge reads 0 4psi 30mbar or more change the filter element Do so even if this occurs bef...

Page 17: ...1059 Paragon Way Rock Hill SC 29730 888 4 MEDGAS 888 463 3427 Fax 803 817 5750 www beaconmed s com...