Medical Vacuum Filter Modules

10.  Install new filter by pressing firmly upwards 

 

 

into the filter head until the filter is seated.

11.  Install  new  filter  housing  O-ring  that  is 

 

 

provided  with  kit  and  re-install  housing  by

 

 

fully tightening the housing onto the head.

12.  Re-install the drain flask by screwing it onto 

 

 

the ball valve located on the bottom of the 

 

 

filter  and  re-open  the  butterfly  valves  on 

 

 

both sides of the filter.

13.  Lastly, dispose of all contaminated PPE.

Summary of Contents for 4107 4018 70

Page 1: ...Installation Operation and Maintenance Instructions Medical Vacuum Filter Modules Part number 4107 9022 25 00 24 July 2018...

Page 2: ......

Page 3: ...l number Serial number Option s included Any information service or spare parts requests should include the machine serial number and be directed to BeaconMed s 1059 Paragon Way Rock Hill SC 29730 Tel...

Page 4: ...Upon Receiving 2 1 2 2 Handling 2 1 2 3 Location 2 1 2 4 Space Requirements 2 1 2 5 Inlet and Outlet Piping 2 1 2 6 Floor Mounting Instructions 2 1 2 7 Drain Flask Installation Instructions 2 1 3 0 O...

Page 5: ...ccor dance with requirements stated by NFPA 99 ISO 7396 1 HTM 02 01 and all applicable local codes 1 5 Environmental Declarations Disposal General When developing products and services Bea conMed s tr...

Page 6: ...contaminated substances and materials separately according to local applicable environ mental legislations Dispose all components according to the appli cable disposal regulations 1 2 Medical Vacuum F...

Page 7: ...ndoors in a clean well ventilated environment Areas of excessive dust dirt or other air borne particulate should be avoided 2 4 Space Requirements The MV Filter Modules should be placed to ensure easy...

Page 8: ...ow decals pictured above may or not be included If so discard They are not required for this application 5 Lastly install the drain flask by screwing it onto the ball valve located on the bottom of th...

Page 9: ...cedure In order to comply with ISO code one filter must be in reserve at all times This means that the but terfly valves on both sides of one filter must be closed at all times 3 4 HTM 02 01 2006 Oper...

Page 10: ...lockages in system See maintenance schedule in section 5 2 See operational procedures in sections 3 2 3 4 for the minimum required filters to be offline Inspect tubes for blockages Reduced vacuum leve...

Page 11: ...ist Please contact BeaconMed s to have the Parts List sent to you or downloadanelectronicversionfromthewebsiteat www beaconmedaes com In addition to the PPE kit below are the recommended gloves to be...

Page 12: ...nded to replace the drain flask with every filter element change The drain flask is available through BeaconMed s The P N for the flask is 4107 6558 95 More information can be found by referencing the...

Page 13: ...lter assembly and ease in removing the filter housing 4 Unscrew the filter housing A strap wrench will ease in loosening the filter housing After the filter housing is loosened it can be unscrewed by...

Page 14: ...ilter the kit P N is 4107 4018 71 More information can be found by referencing the Parts List Please contact BeaconMed s to have the Parts List sent to you or download an electronic version from the w...

Page 15: ...eated 11 Install new filter housing O ring that is provided with kit and re install housing by fully tightening the housing onto the head 12 Re install the drain flask by screwing it onto the ball val...

Page 16: ...6558 95 Replace with every filter element change Differential Pressure Gauge 1629 0438 00 Check monthly If the gauge reads 0 4psi 30mbar or more change the filter element Do so even if this occurs bef...

Page 17: ...1059 Paragon Way Rock Hill SC 29730 888 4 MEDGAS 888 463 3427 Fax 803 817 5750 www beaconmed s com...