2.0 Installation
2.1 Inspection Upon Receiving
The condition of the MV Filter Module should be
carefully inspected upon delivery. Any indication
of damage by the carrier should be noted on
the delivery receipt, especially if the system
will not be immediately uncrated and installed.
The MV Filter Modules may remain in their
shipping containers until ready for installation.
If the MV Filter Modules are to be stored prior
to installation, they must be protected from
the elements to prevent rust and deterioration.
NOTE:
Do not remove the protective covers from the
inlet and discharge connection ports of the
unit until they are ready for connecting to the
hospital’s pipeline distribution system.
2.2 Handling
WARNING:
Use appropriate load rated lifting equipment
and observe safe lifting procedures during all
moves.
2.3 Location
The MV Filter Modules should be installed indoors
in a clean, well-ventilated environment. Areas of
excessive dust, dirt or other air-borne particulate
should be avoided.
2.4 Space Requirements
The MV Filter Modules should be placed to ensure
easy access to perform maintenance and high
visibility of gauges. It is recommended that a
minimum space of 24” be allowed on all sides.
2.5 Inlet and Outlet Piping
All pipe must be either seamless copper tubing
or other corrosion resistant metallic tubing, as
detailed in NFPA 99.
2.6 Floor Mounting Instructions
An Installation Diagram is available as a
supplement to this Operation and Maintenance
Manual. Please contact BeaconMedæs to have
the Installation Diagram sent to you or download
an electronic version from the website at
www.beaconmedaes.com.
2.7 Drain Flask Installation In-
structions
1. Locate the box with the label “ADDITIONAL
PARTS INSIDE”. Contents of the box should
include a cap with integrated fitting,
glass flask, two sealing washers, biohazard
decal, and two arrow decals.
2. Place the biohazard decal onto flask.
2-1
4107 9022 25.00
Medical Vacuum Filter Modules
Summary of Contents for 4107 4018 70