5-1

4107 9022 25.00

Medical Vacuum Filter Modules

1.  Close  butterfly  valves  on  both  sides  of 

 

 

filter being serviced.

5.0  Maintenance 

5.1  Maintenance Procedures

A  PPE  kit  is  available  through  BeaconMedæs

The  kit  includes  a  XL  disposable  coverall, 

N100  disposable  respirator,  wrap-around 

safety  glasses,  and  two  heavy-duty  30-gallon 

biohazard  waste  bags.  The  P/N  for  the  kit  is 

 

4107  4018  65.  More  information  can  be  found 

by  referencing  the  Parts  List.  Please  contact 

BeaconMedæs to have the Parts List sent to you or 

download an electronic version from the website at

 

www.beaconmedaes.com.

In  addition  to  the  PPE  kit,  below  are  the 

recommended gloves to be worn when servicing 

this equipment:

  Grainger P/N: 2VLZ8

 

Description: 9-1/2” Powder Free   

 

 

Unlined Nitrile Disposable Gloves,  

  

Black,Size L, 100PK 

  Grainger P/N: 2VLZ9

 

Description: 9-1/2” Powder Free   

  

Unlined Nitrile Disposable Gloves,  

  

Black, Size XL, 100PK

 WARNING:

  Filter element and drain flask are  

 

biohazard materials and need to be  

 

handled with proper care.

 WARNING:

Proper Personal Protective Equipment (PPE) 

must be worn when servicing this equipment.

Summary of Contents for 4107 4018 70

Page 1: ...Installation Operation and Maintenance Instructions Medical Vacuum Filter Modules Part number 4107 9022 25 00 24 July 2018...

Page 2: ......

Page 3: ...l number Serial number Option s included Any information service or spare parts requests should include the machine serial number and be directed to BeaconMed s 1059 Paragon Way Rock Hill SC 29730 Tel...

Page 4: ...Upon Receiving 2 1 2 2 Handling 2 1 2 3 Location 2 1 2 4 Space Requirements 2 1 2 5 Inlet and Outlet Piping 2 1 2 6 Floor Mounting Instructions 2 1 2 7 Drain Flask Installation Instructions 2 1 3 0 O...

Page 5: ...ccor dance with requirements stated by NFPA 99 ISO 7396 1 HTM 02 01 and all applicable local codes 1 5 Environmental Declarations Disposal General When developing products and services Bea conMed s tr...

Page 6: ...contaminated substances and materials separately according to local applicable environ mental legislations Dispose all components according to the appli cable disposal regulations 1 2 Medical Vacuum F...

Page 7: ...ndoors in a clean well ventilated environment Areas of excessive dust dirt or other air borne particulate should be avoided 2 4 Space Requirements The MV Filter Modules should be placed to ensure easy...

Page 8: ...ow decals pictured above may or not be included If so discard They are not required for this application 5 Lastly install the drain flask by screwing it onto the ball valve located on the bottom of th...

Page 9: ...cedure In order to comply with ISO code one filter must be in reserve at all times This means that the but terfly valves on both sides of one filter must be closed at all times 3 4 HTM 02 01 2006 Oper...

Page 10: ...lockages in system See maintenance schedule in section 5 2 See operational procedures in sections 3 2 3 4 for the minimum required filters to be offline Inspect tubes for blockages Reduced vacuum leve...

Page 11: ...ist Please contact BeaconMed s to have the Parts List sent to you or downloadanelectronicversionfromthewebsiteat www beaconmedaes com In addition to the PPE kit below are the recommended gloves to be...

Page 12: ...nded to replace the drain flask with every filter element change The drain flask is available through BeaconMed s The P N for the flask is 4107 6558 95 More information can be found by referencing the...

Page 13: ...lter assembly and ease in removing the filter housing 4 Unscrew the filter housing A strap wrench will ease in loosening the filter housing After the filter housing is loosened it can be unscrewed by...

Page 14: ...ilter the kit P N is 4107 4018 71 More information can be found by referencing the Parts List Please contact BeaconMed s to have the Parts List sent to you or download an electronic version from the w...

Page 15: ...eated 11 Install new filter housing O ring that is provided with kit and re install housing by fully tightening the housing onto the head 12 Re install the drain flask by screwing it onto the ball val...

Page 16: ...6558 95 Replace with every filter element change Differential Pressure Gauge 1629 0438 00 Check monthly If the gauge reads 0 4psi 30mbar or more change the filter element Do so even if this occurs bef...

Page 17: ...1059 Paragon Way Rock Hill SC 29730 888 4 MEDGAS 888 463 3427 Fax 803 817 5750 www beaconmed s com...

Reviews: