5-6
Medical Vacuum Filter Modules
4107 9022 25.00
5.2 Maintenance Schedule
Item
Kit Part Number
Recommended Maintenance Frequency
MV80 Filter Element
4107 4018 70
Replace annually (*or as needed)
MV160 Filter Element
4107 4018 71
Replace annually (*or as needed)
Drain Flask
4107 6558 95
Replace with every filter element change
Differential Pressure
Gauge
1629 0438 00
Check monthly
*If the gauge reads 0.4psi (≈30mbar) or more, change
the filter element. Do so even if this occurs before a
year has gone by before the filter element was changed
last.
If gauge is damaged, purchase Kit Part Number on the
left.
MV Filter Module
–
Check monthly for blockages and leaks in the system. If
a leak is suspected, pressurize the system to maximum
30 psi and check for leaks.
Summary of Contents for 4107 4018 70