1.0  Introduction 

1.1  Audience

This  manual  provides  information  related  to 

the  installation,  operation,  and  maintenance  of 

the  Medical  Vacuum  (MV)  Filter  Modules  manu

-

factured  by  BeaconMedæs.  Service  information 

contained  in  this  manual  is  intended  for  use  by 

technicians  or  personnel  qualified to  repair  and 

service medical equipment.

1.2 

Definition of Statements

Statements in this manual preceded by following 

words are of special significance.

1.3  Product Description

The MV Filter Modules are designed to be a filter 

for  the  central  vacuum  system  in  hospitals  and 

surgery centers and are located in the plant room 

(or basement) of these facilities.

The filters are designed to protect the vacuum sys

-

tem from liquid, solid, or bacterial contamination. 

There is a transparent drain flask on the bottom 

of each filter that collects liquids and can be eas

-

ily removed for sterilization. The filters are placed 

before the vacuum system. By filtering bacterial 

contamination before the vacuum system, the fil

-

ters not only protect the vacuum system, but also 

the men and women who service the vacuum sys

-

tem. The moving air created by the vacuum sys

-

tem  eventually  gets  exhausted  into  the  outside 

air. The filters also protect people in the commu

-

nity where the exhaust gases are being released 

into the outside air.

1.4  Environmental Considerations

The MV Filter Modules are to be installed in accor

-

dance  with  requirements  stated  by  NFPA  99,  ISO 

7396-1, HTM 02-01, and all applicable local codes.

1.5  Environmental Declarations

Disposal

General

 

When  developing  products  and  services,  Bea

-

conMedæs  tries  to  understand,  address,  and 

minimize the negative environmental effects that 

the products and services may have, when being 

manufactured, distributed, and used, as well as at 

their disposal. 

Recycling  and  disposal  policies  are  part  of  the 

development of all BeaconMedæs products. Bea

-

conMedæs  company  standards  determine  strict 

requirements. 

When selecting materials the substantial recyca

-

bility, the disassembly possibilities and the sepa

-

rability of the materials and assemblies are con

-

sidered  as  well  as  the  environmental  perils  and 

dangers  to  health  during  the  recycling  and  dis

-

posal of the unavoidable rates of non-recycable 

materials. 

BeaconMedæs  products  for  the  most  part  con

-

sist of metallic materials that can be remelted in 

steelworks and smelting works and that is there

-

fore almost infinitely recyclable. The plastic use is 

labeled;  sorting  and  fractioning  of  the  materials 

for recycling in the future is foreseen.

WARNING:

Means there is a possibility of injury or death 

to yourself or others.

 NOTE:

This concept can only succeed with your 

help. 

Support us by disposing professionally. 

By assuring a correct disposal of the product 

you help to prevent possible negative 

consequences for environment and health 

that can occur with inappropriate waste 

handling. 

Recycling and re-usage of materials helps 

to preserve natural resources. 

1-1

4107 9022 25.00

Medical Vacuum Filter Modules

Summary of Contents for 4107 4018 70

Page 1: ...Installation Operation and Maintenance Instructions Medical Vacuum Filter Modules Part number 4107 9022 25 00 24 July 2018...

Page 2: ......

Page 3: ...l number Serial number Option s included Any information service or spare parts requests should include the machine serial number and be directed to BeaconMed s 1059 Paragon Way Rock Hill SC 29730 Tel...

Page 4: ...Upon Receiving 2 1 2 2 Handling 2 1 2 3 Location 2 1 2 4 Space Requirements 2 1 2 5 Inlet and Outlet Piping 2 1 2 6 Floor Mounting Instructions 2 1 2 7 Drain Flask Installation Instructions 2 1 3 0 O...

Page 5: ...ccor dance with requirements stated by NFPA 99 ISO 7396 1 HTM 02 01 and all applicable local codes 1 5 Environmental Declarations Disposal General When developing products and services Bea conMed s tr...

Page 6: ...contaminated substances and materials separately according to local applicable environ mental legislations Dispose all components according to the appli cable disposal regulations 1 2 Medical Vacuum F...

Page 7: ...ndoors in a clean well ventilated environment Areas of excessive dust dirt or other air borne particulate should be avoided 2 4 Space Requirements The MV Filter Modules should be placed to ensure easy...

Page 8: ...ow decals pictured above may or not be included If so discard They are not required for this application 5 Lastly install the drain flask by screwing it onto the ball valve located on the bottom of th...

Page 9: ...cedure In order to comply with ISO code one filter must be in reserve at all times This means that the but terfly valves on both sides of one filter must be closed at all times 3 4 HTM 02 01 2006 Oper...

Page 10: ...lockages in system See maintenance schedule in section 5 2 See operational procedures in sections 3 2 3 4 for the minimum required filters to be offline Inspect tubes for blockages Reduced vacuum leve...

Page 11: ...ist Please contact BeaconMed s to have the Parts List sent to you or downloadanelectronicversionfromthewebsiteat www beaconmedaes com In addition to the PPE kit below are the recommended gloves to be...

Page 12: ...nded to replace the drain flask with every filter element change The drain flask is available through BeaconMed s The P N for the flask is 4107 6558 95 More information can be found by referencing the...

Page 13: ...lter assembly and ease in removing the filter housing 4 Unscrew the filter housing A strap wrench will ease in loosening the filter housing After the filter housing is loosened it can be unscrewed by...

Page 14: ...ilter the kit P N is 4107 4018 71 More information can be found by referencing the Parts List Please contact BeaconMed s to have the Parts List sent to you or download an electronic version from the w...

Page 15: ...eated 11 Install new filter housing O ring that is provided with kit and re install housing by fully tightening the housing onto the head 12 Re install the drain flask by screwing it onto the ball val...

Page 16: ...6558 95 Replace with every filter element change Differential Pressure Gauge 1629 0438 00 Check monthly If the gauge reads 0 4psi 30mbar or more change the filter element Do so even if this occurs bef...