2.0  Installation 

2.1  Inspection Upon Receiving

The condition of the MV Filter Module should be 

carefully inspected upon delivery. Any indication 

of  damage  by  the  carrier  should  be  noted  on 

the  delivery  receipt,  especially  if  the  system 

will  not  be  immediately  uncrated  and  installed. 

The  MV  Filter  Modules  may  remain  in  their 

shipping  containers  until  ready  for  installation. 

If  the  MV  Filter  Modules  are  to  be  stored  prior 

to  installation,  they  must  be  protected  from 

the  elements  to  prevent  rust  and  deterioration.

NOTE:

Do not remove the protective covers from the 

inlet and discharge connection ports of the 

unit until they are ready for connecting to the 

hospital’s pipeline distribution system.

2.2  Handling

WARNING:

Use appropriate load rated lifting equipment 
and observe safe lifting procedures during all 
moves.

2.3  Location

The MV Filter Modules should be installed indoors 

in a clean, well-ventilated environment. Areas of 

excessive dust, dirt or other air-borne particulate 

should be avoided.

2.4  Space Requirements

The MV Filter Modules should be placed to ensure 

easy  access  to  perform  maintenance  and  high 

visibility  of  gauges.  It  is  recommended  that  a 

minimum space of 24” be allowed on all sides.

2.5  Inlet and Outlet Piping

All  pipe  must  be  either  seamless  copper  tubing 

or  other  corrosion  resistant  metallic  tubing,  as 

detailed in NFPA 99.

2.6  Floor Mounting Instructions

An  Installation  Diagram  is  available  as  a 

supplement  to  this  Operation  and  Maintenance 

Manual.  Please  contact  BeaconMedæs  to  have 

the Installation Diagram sent to you or download 

an electronic version from the website at 

www.beaconmedaes.com.

2.7  Drain Flask Installation In-

structions

1.   Locate  the  box  with  the  label  “ADDITIONAL 

 

 

PARTS  INSIDE”.  Contents  of  the  box  should 

 

 

include  a  cap  with  integrated  fitting, 

 

 

glass  flask,  two  sealing  washers,  biohazard 

 

 

decal, and two arrow decals.

2.   Place the biohazard decal onto flask.

2-1

4107 9022 25.00

Medical Vacuum Filter Modules

Summary of Contents for 4107 4018 70

Page 1: ...Installation Operation and Maintenance Instructions Medical Vacuum Filter Modules Part number 4107 9022 25 00 24 July 2018...

Page 2: ......

Page 3: ...l number Serial number Option s included Any information service or spare parts requests should include the machine serial number and be directed to BeaconMed s 1059 Paragon Way Rock Hill SC 29730 Tel...

Page 4: ...Upon Receiving 2 1 2 2 Handling 2 1 2 3 Location 2 1 2 4 Space Requirements 2 1 2 5 Inlet and Outlet Piping 2 1 2 6 Floor Mounting Instructions 2 1 2 7 Drain Flask Installation Instructions 2 1 3 0 O...

Page 5: ...ccor dance with requirements stated by NFPA 99 ISO 7396 1 HTM 02 01 and all applicable local codes 1 5 Environmental Declarations Disposal General When developing products and services Bea conMed s tr...

Page 6: ...contaminated substances and materials separately according to local applicable environ mental legislations Dispose all components according to the appli cable disposal regulations 1 2 Medical Vacuum F...

Page 7: ...ndoors in a clean well ventilated environment Areas of excessive dust dirt or other air borne particulate should be avoided 2 4 Space Requirements The MV Filter Modules should be placed to ensure easy...

Page 8: ...ow decals pictured above may or not be included If so discard They are not required for this application 5 Lastly install the drain flask by screwing it onto the ball valve located on the bottom of th...

Page 9: ...cedure In order to comply with ISO code one filter must be in reserve at all times This means that the but terfly valves on both sides of one filter must be closed at all times 3 4 HTM 02 01 2006 Oper...

Page 10: ...lockages in system See maintenance schedule in section 5 2 See operational procedures in sections 3 2 3 4 for the minimum required filters to be offline Inspect tubes for blockages Reduced vacuum leve...

Page 11: ...ist Please contact BeaconMed s to have the Parts List sent to you or downloadanelectronicversionfromthewebsiteat www beaconmedaes com In addition to the PPE kit below are the recommended gloves to be...

Page 12: ...nded to replace the drain flask with every filter element change The drain flask is available through BeaconMed s The P N for the flask is 4107 6558 95 More information can be found by referencing the...

Page 13: ...lter assembly and ease in removing the filter housing 4 Unscrew the filter housing A strap wrench will ease in loosening the filter housing After the filter housing is loosened it can be unscrewed by...

Page 14: ...ilter the kit P N is 4107 4018 71 More information can be found by referencing the Parts List Please contact BeaconMed s to have the Parts List sent to you or download an electronic version from the w...

Page 15: ...eated 11 Install new filter housing O ring that is provided with kit and re install housing by fully tightening the housing onto the head 12 Re install the drain flask by screwing it onto the ball val...

Page 16: ...6558 95 Replace with every filter element change Differential Pressure Gauge 1629 0438 00 Check monthly If the gauge reads 0 4psi 30mbar or more change the filter element Do so even if this occurs bef...

Page 17: ...1059 Paragon Way Rock Hill SC 29730 888 4 MEDGAS 888 463 3427 Fax 803 817 5750 www beaconmed s com...

Reviews: