5-6
“Oil-Less” Scroll Laboratory Air
4107 9006 25.11
5.0 Maintenance
5.2.8 5 & 10 Hp Only Service Intervals
PART / KIT
NUMBER
QTY
ACTION
AT EACH 5,000 RUNNING HOURS OR
EVERY 2-YEARS
5 Hp
4107 4018 81
1 per unit
Tip Seal Kit
Orbit bearing
Pincrank bearing rotor side
Cooling duct system
Rotor fi ns
Stator fi ns
Replace
Regrease: 3.9 cm3
(6 strokes**)
Regrease: 2 cm3
(3 strokes**)
Clean
Clean
Clean
10 Hp
4107 4018 82
1 per unit
* Intervals:
Cleaning of the parts guiding the cooling air and replacement of the tipseals should be performed
during every service interval.
5.2.9 7.5 & 15 Hp Only Service Intervals
PART / KIT
NUMBER
QTY
ACTION
AT EACH 5,000 RUNNING HOURS OR
EVERY 2-YEARS
7.5 Hp
4107 4000 56
1 per unit
Tip Seal Kit
Orbit bearing
Pincrank bearing rotor side
Cooling duct system
Rotor fi ns
Stator fi ns
Replace
Regrease: 3.9 cm3
(6 strokes**)
Regrease: 2 cm3
(3 strokes**)
Clean
Clean
Clean
15 Hp
4107 4000 58
1 per unit
** Strokes:
Number of strokes valid with recommended purchased grease pump from
Grease Pump Kit 2903 0139 00 and Grease Cartridge 2892 6100 20