C-19
“Oil-Less” Scroll Laboratory Air
4107 9006 25.11
Appendix C: TotalAlert Embedded Control System
4. For each address, select “System” for that
person to receive all alarm/shutdown alerts.
Select “Maintenance” for that person to
receive all maintenance alerts. An individual
may receive both types of alerts.
5. Enter an email address in the “From” box as
this will be the sender of the notifi cations.
6. If desired, early warning notifi cations can be
set up for maintenance items. By selecting
these time intervals, a notifi cation will be sent
before the maintenance item is due, providing
time for planning the service activity.
7. Click the Submit button.
C.7.11 Navigating the Website
The TotalAlert Embedded website allows the user
to easily view the status of all activity pertaining to
the medical air system. By clicking the menu items
to the left of the screen, a user can view pages
displaying accurate and timely information about
the system. These pages include:
Figure C.37 Status Page
Status:
The Status page (Figure C.37)displays
current information for both the overall system
(run hours, pressure, ambient temperature,
alarms, shutdown, and service) and the units (run
hours, operation mode, run mode, sequence,
alarms, shutdown, and service).
Figure C.38 Alarms Page
Alarms:
The Alarms page (Figure C.38) displays
the current status of all system alarms and alarm
status for each unit.
Figure C.39 Shutdown Page
Shutdown:
The Shutdown page (Figure C.39)
displays the current status of all system shutdowns
and shutdown status for each unit.