C-24
“Oil-Less” Scroll Laboratory Air
4107 9006 25.11
Appendix C: TotalAlert Embedded Control System
C.8 BACnet Activation
C.8.1 BACnet Activation
To activate the BACnet device, go to the TotalAlert
Embedded web page for the system, using the
IP address found in the set up menu on the
touchscreen. Enter the password protected Setup
section by logging in with the basic password (see
section C.4).
C.8.2 BACnet Setup
On the BACnet Setup page, set the BACnet Device
ID#. The default value is 10000 but the owner
must set this to the Device Number they want
for their BACnet Network. Enter a name for the
BACnet Device after assigning the ID# (see Figure
C.49).
C.8.3 BACnet Unlock
After submitting the ID# and name, the BACnet
Unlock Screen appears. If system shipped with the
BACnet Activation complete, the Activation code
is populated and the readout states “Unlocked”
(see Figure C.50).
If the system is not unlocked during production,
the Activation code must be entered at this time.
Enter the code and press Submit. When the page
is reloaded, the code will remain in place and the
box to the right will show “Unlocked”.
Figure C.50 BACnet Unlock
Figure C.49 BACnet Setup