Checking the alarm conditions (AGS Simplex)
Isolate and open the panel, and open the MCB (Miniature Circuit Breaker) within the control panel. Switch
a remote switch to ON. Check that after 15 seconds the red PLANT EMERGENCY indicator is illuminated.
Checking the alarm conditions (Duplex installation)
1. Set one pump to HAND, check that the amber PLANT FAULT indicator is illuminated on all remote
switch panels.
2. Isolate and open both panels, and open the three phase MCB (Miniature Circuit Breaker) in each panel.
Close both panels and switch isolators back to ON. Switch a remote switch to ON and check that after 15
seconds, the red PLANT EMERGENCY indicator is illuminated on all remote switch panels. Isolate both
panels and close both three phase MCB's. Close both panels, and switch back to ON.
Endurance test
The complete disposal system must be operated continuously for one hour with a selection of terminal units
in use. Ensure that the system continues to operate within the design flow rate and the exhauster units operate
effectively with no signs of overheating.
Instruction book
30
Summary of Contents for AGS 1560
Page 2: ......
Page 11: ...Simplified schematic of a AGS Simplex installation Instruction book 9...
Page 13: ...AGS Simplex AGS Duplex Instruction book 11...
Page 16: ...Service diagram AGS Simplex 1 ph typical Instruction book 14...
Page 46: ......
Page 47: ......
Page 48: ......