80
4233 5000 13
10 Technical data
10.1 Fuse values
Attention
The indicated fuse value is the maximum value. The cable size used may specify fuses of
a lower value.
Values
Frequency
Voltage
Fuse specification
Maximum fuse value
IEC
50 Hz
230 V
gL/gG
2 A
CSA/UL
60 Hz
115 V
CSA HRC/UL RK5
2 A
CSA/UL
60 Hz
230 V
CSA HRC/UL RK5
2 A
10.2 Electric cable size
Attention
Local regulations remain applicable if they are stricter than the values proposed below.
The voltage drop must not exceed 5% of the nominal voltage. It may be necessary to use
cables of a larger size than those stated to comply with this requirement.
Frequency
Voltage
Minimum cable size
IEC
50 Hz
230 V
3 x 1.5 mm²
CSA/UL
60 Hz
115 V
3 x AWG14
CSA/UL
60 Hz
230 V
3 x AWG14
10.3 Device settings
Regulating and safety devices
The regulating and safety devices are factory-adjusted to obtain optimum performance
of the dryer. Do not alter the setting of any of the devices.