7
Adjustments and servicing procedures
7.1
Servicing the water separator drain
Water separator drain (WSD)
Procedure
Once every six months, service the automatic drain assembly:
1. Isolate the water separator from the air net by closing the external inlet and outlet valve of the unit
(customer’s installation).
2. Depressurize the separator by opening the manual drain valve on connection (4).
3. Unscrew the bowl (2). A whistling noise will warn you if the bowl is not fully depressurized. If this occurs,
the bowl should be screwed back and the venting should be repeated.
4. Remove the drain valve (3) by unscrewing the retaining nut (5) underneath the bowl.
5. Check for clogging of the drain hole. Clean as required. Inspect the drain valve assembly.
6. Reinstall the drain valve in the bowl using the retaining nut.
7. Apply a small amount of water based lubricant (supplied in the medical cleaning kit) on the O-ring and
on the thread of the bowl. Screw the bowl completely on the separator head (1).
8. Repressurize and check for leaks.
Instruction book
72
2920 7099 70
Summary of Contents for dMED 100
Page 2: ......
Page 40: ...Inputs Instruction book 38 2920 7099 70...
Page 45: ...Instruction book 2920 7099 70 43...
Page 54: ...4 Installation 4 1 Dimension drawings Instruction book 52 2920 7099 70...
Page 55: ...Instruction book 2920 7099 70 53...
Page 56: ...Instruction book 54 2920 7099 70...
Page 57: ...Instruction book 2920 7099 70 55...
Page 58: ...Instruction book 56 2920 7099 70...
Page 59: ...Instruction book 2920 7099 70 57...
Page 60: ...Instruction book 58 2920 7099 70...
Page 61: ...Instruction book 2920 7099 70 59...
Page 62: ...References on drawings Instruction book 60 2920 7099 70...
Page 93: ......
Page 94: ...No 2920 7099 70 2012 01 Printed in Belgium www beaconmedaes com...