Condition
Fault
Remedy
Pressure sensors: error signal
(no measurement)
•
Electric connection
interrupted
•
Mechanical damage (e.g.
too high mounting torque)
•
Connection plugs not
connected
•
Crack due to incorrect
mounting
•
Internal defect
•
No supply voltage
•
Check connections
•
Check sensors
Pressure sensors: too high or
too low indication
•
Condense on connection
plugs
•
Faulty supply signal
•
Clogging or contamination
•
Sensor internally defect
•
Deviation due to ageing
•
Deviation due to incorrect
production
•
Check sensors
•
Replace defective parts
Pressure regulator: wrong
outlet pressure
Wrong setting
•
Check gauge of pressure regulator
•
Adjust pressure
Dew point sensor
malfunctioning
No direct feedback of dew point Replace defective parts
Dew point sensor: no
measurement (error signal)
No direct feedback of dew point
•
Check connections
•
Check sensors
Dew point sensor: incorrect
measurement
•
Dew point alarm generated
too early or too late
•
Check sensors
•
Replace defective parts
Faults and remedies, ES-MED
Condition
Fault
Remedy
Doesn't read in pressures
Software fault
Reprogram the ES-MED
Doesn't read in (correct)
compressor status
Software fault
Reprogram the ES-MED
Doesn't produce run queue
Software fault
Reprogram the ES-MED
Doesn't operate the
compressors
•
Software fault
•
Communication fault
•
Reprogram the ES-MED
•
Check wiring
Gives out <Run> or <Not run>
command too slowly
Software fault
Reprogram the ES-MED
Doen't issue alarm(s) when it
should
•
Software fault
•
Connection fault between
ES-MED and hospital
control room display
•
Reprogram the ES-MED
•
Check wiring
Doesn't issue service warnings
when it should
Software fault
Reprogram the ES-MED
Keeps exhaust (or inlet) valves
closed (or opened)
•
Software fault
•
Communication fault
•
Reprogram the ES-MED
•
Check wiring
•
Check solenoid
Instruction book
2920 7099 70
79
Summary of Contents for dMED 100
Page 2: ......
Page 40: ...Inputs Instruction book 38 2920 7099 70...
Page 45: ...Instruction book 2920 7099 70 43...
Page 54: ...4 Installation 4 1 Dimension drawings Instruction book 52 2920 7099 70...
Page 55: ...Instruction book 2920 7099 70 53...
Page 56: ...Instruction book 54 2920 7099 70...
Page 57: ...Instruction book 2920 7099 70 55...
Page 58: ...Instruction book 56 2920 7099 70...
Page 59: ...Instruction book 2920 7099 70 57...
Page 60: ...Instruction book 58 2920 7099 70...
Page 61: ...Instruction book 2920 7099 70 59...
Page 62: ...References on drawings Instruction book 60 2920 7099 70...
Page 93: ......
Page 94: ...No 2920 7099 70 2012 01 Printed in Belgium www beaconmedaes com...