9
Technical data
9.1
Fuse values
Attention
The indicated fuse value is the maximum value. The cable size used may specify fuses of
a lower value.
Values
Frequency
Voltage
Fuse specification
Maximum fuse value
IEC
50 Hz
230 V
gL/gG
2 A
CSA/UL
60 Hz
115 V
CSA HRC/UL RK5
2 A
CSA/UL
60 Hz
230 V
CSA HRC/UL RK5
2 A
9.2
Electric cable size
Attention
Local regulations remain applicable if they are stricter than the values proposed below.
The voltage drop must not exceed 5% of the nominal voltage. It may be necessary to use
cables of a larger size than those stated to comply with this requirement.
Frequency
Voltage
Minimum cable size
IEC
50 Hz
230 V
3 x 1.5 mm²
CSA/UL
60 Hz
115 V
3 x AWG14
CSA/UL
60 Hz
230 V
3 x AWG14
9.3
Device settings
Regulating and safety devices
The regulating and safety devices are factory-adjusted to obtain optimum performance of the dryer. Do not
alter the setting of any of the devices.
Instruction book
80
2920 7099 70
Summary of Contents for dMED 100
Page 2: ......
Page 40: ...Inputs Instruction book 38 2920 7099 70...
Page 45: ...Instruction book 2920 7099 70 43...
Page 54: ...4 Installation 4 1 Dimension drawings Instruction book 52 2920 7099 70...
Page 55: ...Instruction book 2920 7099 70 53...
Page 56: ...Instruction book 54 2920 7099 70...
Page 57: ...Instruction book 2920 7099 70 55...
Page 58: ...Instruction book 56 2920 7099 70...
Page 59: ...Instruction book 2920 7099 70 57...
Page 60: ...Instruction book 58 2920 7099 70...
Page 61: ...Instruction book 2920 7099 70 59...
Page 62: ...References on drawings Instruction book 60 2920 7099 70...
Page 93: ......
Page 94: ...No 2920 7099 70 2012 01 Printed in Belgium www beaconmedaes com...