8. Click on the
Floor
window.
9. Type in the building floor that this Master Alarm is located on.
NOTE:
Available levels are -2, which is a sub-basement level, through 100.
10. Click on the
Direction
pull down menu arrow (see figure 1-37 on page
1–87).
Figure 1-37. Setup Device—Wing Direction
11. Click on the direction of the wing that this Master Alarm is located on.
12. Click on the
Silenced return time
pull down menu arrow (see figure 1-38 on
page 1–88).
MedPlus TotalAlert® Alarm Network Operation and Maintenance Manual (205728)
Page 1-87
Configuring and Browsing the Master Alarm Using the Web Pages
Chapter 1: Introduction