MedPlus TotalAlert® Alarm Network Operation and Maintenance Manual (205728)
Configuring and Browsing the Master Alarm Using the Web Pages
Chapter 1: Introduction
Configure Electronic Notification
When there is an alarm condition, up to five electronic messages can be sent
from each Master Alarm.
1. To configure the electronic notification, click on the Electronic Notification
link. The Electronic Notification page displays (see figure 1-40 on page
1–90).
Figure 1-40. Electronic Notification
2. Learn either the
SMTP Server Name
or
SMTP Server Address
.
NOTE:
The facility’s Information System personnel may need to be contacted for the
server name and address that the facility uses for the SMTP server.
NOTE:
The SMTP Server Address has priority over the SMTP Server Name.
3. To use the
SMTP Server Name
, leave the
SMTP Server Address
window
blank.
NOTE:
The SMTP Server Name requires a Domain Name Server (DNS) look-up. Alarms
that occur before the server name is resolved are not sent.
4. If you know the
SMTP Server Address
, click in the
SMTP Server Address
window.