Clear Alarm Network Device List
1. To clear the alarm network device list, click on the Clear Network link. The
clear network page displays (see figure 1-44 on page 1–95).
NOTE:
This command clears the device list memory from all of the devices on the alarm
network. Immediately after the memory is cleared, the devices automatically re-
learn the device list.
Figure 1-44. Clear Network
2. Use it when you add, remove, or exchange the Master Alarm Module,
Network Communications Boards, or the Area Alarm Modules. Otherwise,
alarms will occur when the network does not report as expected.
3. To clear the alarm network’s device list, click on the Click Here link.
MedPlus TotalAlert® Alarm Network Operation and Maintenance Manual (205728)
Page 1-95
Configuring and Browsing the Master Alarm Using the Web Pages
Chapter 1: Introduction