Page 1-96
MedPlus TotalAlert® Alarm Network Operation and Maintenance Manual (205728)
Configuring and Browsing the Master Alarm Using the Web Pages
Chapter 1: Introduction
Transfer Configuration
After one Master Alarm is configured, the following information can be trans-
ferred to another Master Alarm across the network:
• Input Information from each configured input
•
Gas-type
•
Message
•
LED configuration
•
System number
• Silenced return time
NOTE:
To comply with the requirements of the City of New York Department of
Buildings; Materials and Equipment Acceptance Division (M.E.A.), do not set
the alarm return time to “never.” Set the alarm to return every 30 minutes.
• SMTP address
• SMTP server
• User names
• User passwords
NOTE:
The electronic notification recipients and the alarm network settings do not trans-
fer.
1. To transfer the device configuration, click on the Transfer Setup link. The
Configuration Transfer page displays (see figure 1-45 on page 1–97).