Table 1-30. Alarm Network Statistics Descriptions
Network Statistic
Description
IP Address
The IP address is in use.
Sub-net
The sub-net is in use.
Gateway
The router address that is on the sub-net.
MAC Address
The device’s Ethernet address.
Receives
Ethernet packets received.
Unicasts
Ethernet packets that are directed to this device.
Multicasts
Not used.
Broadcasts
Ethernet packets that are sent to all devices.
Rx Errors
Total number of Ethernet receive errors.
Rx missed
Packets not received from the MAC queue in time.
RX CRC errors
Checksum error on a received packet.
Rx drops
Received packets that were lost because there was no buffer space.
Transmits
Ethernet packets that the device has sent.
Buffer defers
Packets that are held off because the buffer is full.
Tx errors
Total number of Ethernet transmit errors.
Tx collisions
A packet that was sent and then collided with another packet.
Tx Coll. overflow
Packets that failed to re-send, after they had collided.
Tx FILO errors
The MAC queue over-filled.
Traffic backoffs
The amount of time the MAC delayed transmitting, due to high traf-
fic.
Network rejects
Packets rejected, because they were not an IP or ARP.
Address rejects
Packet rejected, because they had the wrong address.
Protocol rejects
Packets rejected, because their protocol was not supported.
ARP rejects
Broadcast ARPs rejected, because they were not for this IP.
SMPT server IP
The address of the mail server.
Peer no memory
Not enough memory to build a MedPlus TotalAlert® packet.
Peer ARP waiting
Transmission is delayed until the ARP resolves the MAC address.
Peer no send
MedPlus TotalAlert® packet not sent, because there are no packet buffers.
MedPlus TotalAlert® Alarm Network Operation and Maintenance Manual (205728)
Page 1-105
Configuring and Browsing the Master Alarm Using the Web Pages
Chapter 1: Introduction
1
Summary of Contents for MedPlus TotalAlert
Page 370: ...Schematic Wiring Diagram Typical Local Alarm...
Page 371: ...Schematic Wiring Diagram Typical Master Alarm...
Page 372: ...Schematic Wiring Diagram Typical Master Alarm to Master Alarm...
Page 373: ...Schematic Wiring Diagram Typical Area Alarm...
Page 374: ...Schematic Wiring Diagram Typical Master Alarm to Three Area Alarm...
Page 375: ...Schematic Wiring Diagram Typical Master Alarm to Six Area Alarm Combination...