Page 1-106
MedPlus TotalAlert® Alarm Network Operation and Maintenance Manual (205728)
Configuring and Browsing the Master Alarm Using the Web Pages
Chapter 1: Introduction
Network Statistic
Description
Send queue length Packets that are in the transmit queue.
Send queue max
Packets that are in the transmit queue.
Send queue min.
Packets that are in the transmit queue.
Recv queue length Packets that are in the receive queue.
Recv queue max
Packets that are in the receive queue.
Recv queue min
Packets that are in the receive queue.
Big queue length
Large size packets are available.
Big queue max
Large size packets are available.
Big queue min
Large size packets are available.
Little queue length Small size packets are available.
Little queue max
Small size packets are available.
Little queue min
Small size packets are available.
PMP Rx
RS-485 packets were received.
PMP Tx
RS-485 packets were sent.
PMP bad CRC
RS-485 packets were received with bad checksum.
PMP missed end
RS-485 packets that did not frame correctly.
PMP too long
RS-485 packets that exceeded the expected length.
Area bus timeouts RS-485 requests that had no response.
Summary of Contents for MedPlus TotalAlert
Page 370: ...Schematic Wiring Diagram Typical Local Alarm...
Page 371: ...Schematic Wiring Diagram Typical Master Alarm...
Page 372: ...Schematic Wiring Diagram Typical Master Alarm to Master Alarm...
Page 373: ...Schematic Wiring Diagram Typical Area Alarm...
Page 374: ...Schematic Wiring Diagram Typical Master Alarm to Three Area Alarm...
Page 375: ...Schematic Wiring Diagram Typical Master Alarm to Six Area Alarm Combination...