Browse Input States
1. To browse the input states, click on the Physical Inputs link. The Input
Diagnostics page displays (see figure 1-53 on page 1–107).
Figure 1-53. Input Diagnostics
NOTE:
This page is useful when troubleshooting the alarm system’s wiring.
2. Click on the Refresh link to view the most recent information.
MedPlus TotalAlert® Alarm Network Operation and Maintenance Manual (205728)
Page 1-107
Configuring and Browsing the Master Alarm Using the Web Pages
Chapter 1: Introduction
1
Summary of Contents for MedPlus TotalAlert
Page 370: ...Schematic Wiring Diagram Typical Local Alarm...
Page 371: ...Schematic Wiring Diagram Typical Master Alarm...
Page 372: ...Schematic Wiring Diagram Typical Master Alarm to Master Alarm...
Page 373: ...Schematic Wiring Diagram Typical Area Alarm...
Page 374: ...Schematic Wiring Diagram Typical Master Alarm to Three Area Alarm...
Page 375: ...Schematic Wiring Diagram Typical Master Alarm to Six Area Alarm Combination...