5. If the Network Communications Board is installed, check the three LEDs (see
figure 1-56 on page 1–113).
Figure 1-56. Network Communications Board
a. The first LED (CZ) indicates whether the software is operating correctly.
–
Steady flash indicates proper operation.
–
Solid ON or solid OFF indicates that the module has failed.
–
Rapid flashing indicates a software error.
MedPlus TotalAlert® Alarm Network Operation and Maintenance Manual (205728)
Page 1-113
Area Alarm Power-Up
Chapter 1: Introduction
1