2.1
Transducer Pressure Below Sensor Range
1. Error code
A 01
displays on an Area Alarm Module.
Yes
No
L J
Go to “Final Actions” on page 2-7.
2. The shutoff valve is open.
Yes
No
L J
Open the shutoff valve. If this solves the problem, go to “Final
Actions” on page 2-7. Otherwise, go to step 3.
3. Transducer is connected to the pipeline.
Yes
No
L J
Connect the transducer to the pipeline (refer to procedure 4.20
on page 4-50). If this solves the problem, go to “Final Actions”
on page 2-7. Otherwise, go to step 4.
4. Transducer has a slow steady flash on its heartbeat LED.
Yes
No
L J
Replace the transducer (refer to procedure 4.23 on page 4-62). If
this solves the problem, go to “Final Actions” on page 2-7.
Otherwise, go to step 5.
5. If the problem still exists, call BeaconMedæs Technical Support.
MedPlus TotalAlert® Alarm Network Operation and Maintenance Manual (205728)
Page 2-13
Transducer Pressure Below Sensor Range
Chapter 2: Troubleshooting Procedures
2