Page 2-16
MedPlus TotalAlert® Alarm Network Operation and Maintenance Manual (205728)
Incorrect Transducer
Chapter 2: Troubleshooting Procedures
2.5
Incorrect Transducer
1. An Area Alarm Module displays error code
A 07
.
Yes
No
L J
Go to “Final Actions” on page 2-7.
2. Transducer is programmed for the correct type of gas.
Yes
No
L J
Connect the proper gas-specific transducer (refer to procedure
4.23 on page 4-62). If you need more information about how to
connect the transducer, refer to the
MedPlus TotalAlert® Alarm
Network—Conversion Kits Installation Instructions
. If this
solves the problem, go to “Final Actions” on page 2-7.
Otherwise, go to step 3.
3. Transducer has a slow steady flash on its heartbeat LED.
Yes
No
L J
Replace the transducer (refer to procedure 4.23 on page 4-62). If
this solves the problem, go to “Final Actions” on page 2-7.
Otherwise, go to step 4.
4. If the problem still exists, call BeaconMedæs Technical Support.
2.6
Transducer Program Errors
1. An Area Alarm Module displays error code
A 10
.
Yes
No
L J
Go to “Final Actions” on page 2-7.
2. Transducer has a slow steady flash on its heartbeat LED.
Yes
No
L J
Replace the transducer (refer to procedure 4.23 on page 4-62). If
this solves the problem, go to “Final Actions” on page 2-7.
Otherwise, go to step 3.
3. If the problem still exists, call BeaconMedæs Technical Support.
Summary of Contents for MedPlus TotalAlert
Page 370: ...Schematic Wiring Diagram Typical Local Alarm...
Page 371: ...Schematic Wiring Diagram Typical Master Alarm...
Page 372: ...Schematic Wiring Diagram Typical Master Alarm to Master Alarm...
Page 373: ...Schematic Wiring Diagram Typical Area Alarm...
Page 374: ...Schematic Wiring Diagram Typical Master Alarm to Three Area Alarm...
Page 375: ...Schematic Wiring Diagram Typical Master Alarm to Six Area Alarm Combination...