Page 2-18
MedPlus TotalAlert® Alarm Network Operation and Maintenance Manual (205728)
Area Alarm Program Errors
Chapter 2: Troubleshooting Procedures
2.8
Area Alarm Program Errors
1. An Area Alarm Module displays error code
A 08
or
A 09
.
Yes
No
L J
Go to “Final Actions” on page 2-7.
2. Replace the Area Alarm Module (refer to procedure 4.9 on page 4-26). This
solves the problem.
Yes
No
L J
Call BeaconMedæs Technical Support.
3. Go to “Final Actions” on page 2-7.
2.9
Area Alarm Module RAM Error
1. An Area Alarm Module displays error code
A 04
.
Yes
No
L J
Go to “Final Actions” on page 2-7.
2. Area Alarm Module is connected properly.
Yes
No
L J
Connect the Area Alarm Module properly (refer to procedure
4.9 on page 4-26). If this solves the problem, go to “Final
Actions” on page 2-7. Otherwise, go to step 3.
3. Area Alarm Module is working correctly.
Yes
No
L J
Replace the Area Alarm Module (refer to procedure 4.9 on page
4-26). If this solves the problem, go to “Final Actions” on page
2-7. Otherwise, go to step 4.
4. If the problem still exists, call BeaconMedæs Technical Support.