Page 2-20
MedPlus TotalAlert® Alarm Network Operation and Maintenance Manual (205728)
Ethernet Connection Is Lost
Chapter 2: Troubleshooting Procedures
2.12 Ethernet Connection Is Lost
1. Ethernet Lost
displays on the Master Alarm.
Yes
No
L J
Go to “Final Actions” on page 2-7.
2. The Ethernet is connected to the Master Alarm Module.
Yes
No
L J
Connect the Ethernet to the Master Alarm Module (refer to pro-
cedure 4.25 on page 4-68). If this solves the problem, go to
“Final Actions” on page 2-7. Otherwise, go to step 3.
3. The Ethernet is connected to the corresponding hub.
Yes
No
L J
Connect the Ethernet to the corresponding hub. Refer to the
wiring diagram of the facility’s Ethernet network. If this solves
the problem, go to “Final Actions” on page 2-7. Otherwise, go
to step 4.
4. The Master Alarm Module is operating properly.
Yes
No
L J
Replace the Master Alarm Module (refer to procedure 4.2 on
page 4-10). If this solves the problem, go to “Final Actions” on
page 2-7. Otherwise, call BeaconMedæs Technical Support .
5. Facility’s Ethernet is operating properly.
Yes
No
L J
Call the facility’s Information Systems department. If this solves
the problem, go to “Final Actions” on page 2-7. Otherwise, go
to step 6.
6. If the problem still exists, call BeaconMedæs Technical Support.
Summary of Contents for MedPlus TotalAlert
Page 370: ...Schematic Wiring Diagram Typical Local Alarm...
Page 371: ...Schematic Wiring Diagram Typical Master Alarm...
Page 372: ...Schematic Wiring Diagram Typical Master Alarm to Master Alarm...
Page 373: ...Schematic Wiring Diagram Typical Area Alarm...
Page 374: ...Schematic Wiring Diagram Typical Master Alarm to Three Area Alarm...
Page 375: ...Schematic Wiring Diagram Typical Master Alarm to Six Area Alarm Combination...