2.13 Area Display/Transducer Fault
1. Area Display/Transducer Fault
displays on the Master Alarm.
Yes
No
L J
Go to “Final Actions” on page 2-7.
2. The Area Alarm Module is connected to the Master Alarm.
Yes
No
L J
Connect the Area Alarm to the Master Alarm (refer to procedure
4.9 on page 4-26). If this solves the problem, go to “Final
Actions” on page 2-7. Otherwise, go to step 3.
3. The transducer is connected to the Area Alarm Breakout P.C. Board.
Yes
No
L J
Connect the transducer to the Area Alarm Breakout P.C. Board
(refer to procedure 4.7 on page 4-22). If this solves the problem,
go to “Final Actions” on page 2-7. Otherwise, go to step 4.
4. The Area Alarm Module is operating properly.
Yes
No
L J
Replace the Area Alarm Module (refer to procedure 4.9 on page
4-26). If this solves the problem, go to “Final Actions” on page
2-7. Otherwise, go to step 5.
5. The transducer’s LED is flashing at a slow steady rate.
Yes
No
L J
Replace the transducer (refer to procedure 4.20 on page 4-50). If
this solves the problem, go to “Final Actions” on page 2-7.
Otherwise, go to step 6.
6. If the problem still exists, call BeaconMedæs Technical Support.
MedPlus TotalAlert® Alarm Network Operation and Maintenance Manual (205728)
Page 2-21
Area Display/Transducer Fault
Chapter 2: Troubleshooting Procedures
2
Summary of Contents for MedPlus TotalAlert
Page 370: ...Schematic Wiring Diagram Typical Local Alarm...
Page 371: ...Schematic Wiring Diagram Typical Master Alarm...
Page 372: ...Schematic Wiring Diagram Typical Master Alarm to Master Alarm...
Page 373: ...Schematic Wiring Diagram Typical Area Alarm...
Page 374: ...Schematic Wiring Diagram Typical Master Alarm to Three Area Alarm...
Page 375: ...Schematic Wiring Diagram Typical Master Alarm to Six Area Alarm Combination...