Page 2-22
MedPlus TotalAlert® Alarm Network Operation and Maintenance Manual (205728)
Service Required
Chapter 2: Troubleshooting Procedures
2.14 Service Required
1. Service Required
displays on the Master Alarm.
Yes
No
L J
Go to “Final Actions” on page 2-7.
2. Master Alarm is connected to the source equipment properly.
Yes
No
L J
Connect the source equipment to the Master Alarm. Refer to
either the
MedPlus TotalAlert® Alarm Network Installation
Instructions
or the
MedPlus TotalAlert® Alarm Network—
Conversion Kits Installation Instructions
, if needed. If this
solves the problem, go to “Final Actions” on page 2-7.
Otherwise, go to step 3.
3. If the problem still exists, call BeaconMedæs Technical Support.
2.15 Gas or Vacuum Wrong Gas
1. Wrong Gas
displays on the Master Alarm.
Yes
No
L J
Go to “Final Actions” on page 2-7.
2. The Area Alarm Module, at the location showing on the Master Alarm’s dis-
play, is connected to the correct gas-specific transducer.
Yes
No
L J
Replace the transducer with one that is for the correct type of
gas or vacuum (refer to procedure 4.20 on page 4-50). If this
solves the problem, go to “Final Actions” on page 2-7.
Otherwise, go to step 3.
3. Replace the Area Alarm Module with one that is for the correct type of gas or
vacuum of the line that it is monitoring (refer to procedure 4.9 on page 4-26).
This solves the problem.
Yes
No
L J
Call BeaconMedæs Technical Support.
4. Go to “Final Actions” on page 2-7.
Summary of Contents for MedPlus TotalAlert
Page 370: ...Schematic Wiring Diagram Typical Local Alarm...
Page 371: ...Schematic Wiring Diagram Typical Master Alarm...
Page 372: ...Schematic Wiring Diagram Typical Master Alarm to Master Alarm...
Page 373: ...Schematic Wiring Diagram Typical Area Alarm...
Page 374: ...Schematic Wiring Diagram Typical Master Alarm to Three Area Alarm...
Page 375: ...Schematic Wiring Diagram Typical Master Alarm to Six Area Alarm Combination...