2.16 Gas or Vacuum Wiring
1. A type of
gas
or
vacuum
and the word
Wiring
displays on the Master Alarm.
Yes
No
L J
Go to “Final Actions” on page 2-7.
2. The Area Alarm Module, at the location showing on the Master Alarm’s dis-
play, is connected to its corresponding transducer properly.
Yes
No
L J
Properly connect the Area Alarm Module to its corresponding
transducer (refer to procedure 4.9 on page 4-26) and (refer to
procedure 4.23 on page 4-62). If this solves the problem, go to
“Final Actions” on page 2-7. Otherwise, go to step 3.
3. The wiring between the Area Alarm Module and the transducer is pinched.
Yes
No
L J
Call BeaconMedæs Technical Support.
4. Replace the wiring. If this solves the problem go to “Final Actions” on page
2-7. Otherwise, call BeaconMedæs Technical Support.
2.17 Gas or Vacuum Shorted
1. A type of
gas
or
vacuum
and the word
Shorted
displays on the Master
Alarm.
Yes
No
L J
Go to “Final Actions” on page 2-7.
2. The Area Alarm Module, at the location showing on the Master Alarm’s dis-
play, is connected to its corresponding transducer properly.
Yes
No
L J
Properly connect the Area Alarm Module to its corresponding
transducer (refer to procedure 4.9 on page 4-26 and procedure
4.23 on page 4-62). If this solves the problem, go to “Final
Actions” on page 2-7. Otherwise, go to step 3.
MedPlus TotalAlert® Alarm Network Operation and Maintenance Manual (205728)
Page 2-23
Gas or Vacuum Wiring
Chapter 2: Troubleshooting Procedures
2
Summary of Contents for MedPlus TotalAlert
Page 370: ...Schematic Wiring Diagram Typical Local Alarm...
Page 371: ...Schematic Wiring Diagram Typical Master Alarm...
Page 372: ...Schematic Wiring Diagram Typical Master Alarm to Master Alarm...
Page 373: ...Schematic Wiring Diagram Typical Area Alarm...
Page 374: ...Schematic Wiring Diagram Typical Master Alarm to Three Area Alarm...
Page 375: ...Schematic Wiring Diagram Typical Master Alarm to Six Area Alarm Combination...