6. The Area Alarm’s Network Communications Board is connected to its
Ethernet hub properly.
Yes
No
L J
Properly connect the Area Alarm’s Network Communications
Board to its corresponding Ethernet hub. Refer to the facility’s
wiring diagram of the Ethernet network. If this solves the prob-
lem, go to “Final Actions” on page 2-7. Otherwise, go to step 7.
7. Replace the Area Alarm’s Network Communications Board (refer to proce-
dure 4.8 on page 4-24). If this solves the problem, go to “Final Actions” on
page 2-7. Otherwise, call BeaconMedæs Technical Support.
2.19 Serial Number Lost (Master Alarm)
1. A Network Communications Board serial number and the word
lost
displays
on the
Master Alarm
.
Yes
No
L J
Go to “Final Actions” on page 2-7.
2. The Network Communications Board was replaced or removed from the
alarm network.
Yes
No
L J
Ensure all of the Ethernet connections are connected properly.
Refer to the facility’s wiring diagram of the Ethernet network. If
this solves the problem, go to “Final Actions” on page 2-7.
Otherwise, call BeaconMedæs Technical Support.
3. Do a
Clear Network
command on the Master Alarm; refer to “Configuring
the Master Alarm by Using the Front Panel” on page 1-63, or “Clear Alarm
Network Device List” on page 1-95. This solves the problem.
Yes
No
L J
Call BeaconMedæs Technical Support.
4. Go to “Final Actions” on page 2-7.
MedPlus TotalAlert® Alarm Network Operation and Maintenance Manual (205728)
Page 2-25
Serial Number Lost (Master Alarm)
Chapter 2: Troubleshooting Procedures
2
Summary of Contents for MedPlus TotalAlert
Page 370: ...Schematic Wiring Diagram Typical Local Alarm...
Page 371: ...Schematic Wiring Diagram Typical Master Alarm...
Page 372: ...Schematic Wiring Diagram Typical Master Alarm to Master Alarm...
Page 373: ...Schematic Wiring Diagram Typical Area Alarm...
Page 374: ...Schematic Wiring Diagram Typical Master Alarm to Three Area Alarm...
Page 375: ...Schematic Wiring Diagram Typical Master Alarm to Six Area Alarm Combination...