5. The facility’s supply voltage is connected to the alarm’s terminal block.
Yes
No
L J
Connect the facility’s supply to the alarm’s terminal block. For
the wiring diagram, refer to FO 3-4 at the rear of this document.
If this solves the problem, go to “Final Actions” on page 2-7.
Otherwise, go to step 6.
6. The wiring on the transformer is correct.
Yes
No
L J
Wire the transformer correctly. For the wiring diagram, refer to
FO 3-4 at the rear of this document. If this solves the problem,
go to “Final Actions” on page 2-7. Otherwise, go to step 7.
7. The voltage across the two orange lead connections on the breakout board is
approximately 20 V AC.
Yes
No
L J
Replace the transformer (refer to procedure 4.5 on page 4-16). If
this solves the problem, go to “Final Actions” on page 2-7.
Otherwise, call BeaconMedæs Technical Support.
8. Go to “Final Actions” on page 2-7.
MedPlus TotalAlert® Alarm Network Operation and Maintenance Manual (205728)
Page 2-31
No LEDs or Displays Are Lit on the Area Alarm
Chapter 2: Troubleshooting Procedures
2