2.29 No Audible Signal on the Master Alarm
1. When an alarm occurs, the audible indicator works correctly.
Yes
No
L J
Replace the Master Module (refer to procedure 4.2 on page 4-
10). If this solves the problem, go to “Final Actions” on page 2-
7. Otherwise, call BeaconMedæs Technical Support.
2. Go to “Final Actions” on page 2-7.
2.30 No Audible Signal on the Area Alarm
1. When an alarm occurs, the audible indicator works correctly.
Yes
No
L J
Replace the Area Alarm P.C. board located on the back of the
front panel (refer to procedure 4.8 on page 4-24). If this solves
the problem, go to “Final Actions” on page 2-7. Otherwise, call
BeaconMedæs Technical Support.
2.
Go to “Final Actions” on page 2-7.