Page 2-38
MedPlus TotalAlert® Alarm Network Operation and Maintenance Manual (205728)
No Audible Signal on the Local Alarm
Chapter 2: Troubleshooting Procedures
2.31 No Audible Signal on the Local Alarm
1. The Local Alarm audible indicator does not work.
Yes
No
L J
Go to “Final Actions” on page 2-7.
2. The wiring from the Local Alarm P.C. board on the back of the front panel to
the buzzer is connected.
Yes
No
L J
Connect the buzzer’s wiring (refer to procedure 4.14 on page 4-
38). If this solves the problem, go to “Final Actions” on page 2-
7. Otherwise, go to step 3.
3. The buzzer works properly.
Yes
No
L J
Replace the buzzer. If this solves the problem, go to “Final
Actions” on page 2-7. Otherwise, call BeaconMedæs Technical
Support.
4. Go to “Final Actions” on page 2-7.
2.32 Input # Not Assigned
1. Input # Not Assigned
, displays on the
Master Alarm
.
Yes
No
L J
Go to “Final Actions” on page 2-7.
2. The input number shown on the display is configured.
Yes
No
L J
Configure the input. See “Configuring the Master Alarm by
Using the Front Panel” on page 1-63. If this solves the problem,
go to “Final Actions” on page 2-7. Otherwise, call
BeaconMedæs Technical Support.
3. Call BeaconMedæs Technical Support.
Summary of Contents for MedPlus TotalAlert
Page 370: ...Schematic Wiring Diagram Typical Local Alarm...
Page 371: ...Schematic Wiring Diagram Typical Master Alarm...
Page 372: ...Schematic Wiring Diagram Typical Master Alarm to Master Alarm...
Page 373: ...Schematic Wiring Diagram Typical Area Alarm...
Page 374: ...Schematic Wiring Diagram Typical Master Alarm to Three Area Alarm...
Page 375: ...Schematic Wiring Diagram Typical Master Alarm to Six Area Alarm Combination...