2.33 Master Alarm Cannot be Browsed
1. The Master Alarm cannot be browsed.
Yes
No
L J
Go to “Final Actions” on page 2-7.
2. Press the
Test key
on the front panel. The address that shows on the display
is a series: 169.254. ##. ##.
Yes
No
L J
Ensure that the Ethernet connections are properly connected on
all of the alarm devices (refer to procedure 4.25 on page 4-68).
They are properly connected. If this solves the problem, go to
“Final Actions” on page 2-7. Otherwise, go to step 3.
3. The alarm is not finding a server. Type the word
Master
followed by
its
serial
number into your Internet browser’s address line: for example
Master47856
.
You still cannot browse the Master Alarm.
Yes
No
L J
Go to “Final Actions” on page 2-7.
4. Check the LED that indicates whether the alarms are communicating properly
on the Master Alarm Module. See “Master Alarm Power-Up” on page 1-60.
The LED is flashing at a steady rate.
Yes
No
L J
Call BeaconMedæs Technical Support.
5. If you still cannot browse the Master Alarm, call BeaconMedæs Technical
Support.
MedPlus TotalAlert® Alarm Network Operation and Maintenance Manual (205728)
Page 2-39
Master Alarm Cannot be Browsed
Chapter 2: Troubleshooting Procedures
2
Summary of Contents for MedPlus TotalAlert
Page 370: ...Schematic Wiring Diagram Typical Local Alarm...
Page 371: ...Schematic Wiring Diagram Typical Master Alarm...
Page 372: ...Schematic Wiring Diagram Typical Master Alarm to Master Alarm...
Page 373: ...Schematic Wiring Diagram Typical Area Alarm...
Page 374: ...Schematic Wiring Diagram Typical Master Alarm to Three Area Alarm...
Page 375: ...Schematic Wiring Diagram Typical Master Alarm to Six Area Alarm Combination...