Page 2-40
MedPlus TotalAlert® Alarm Network Operation and Maintenance Manual (205728)
Intermittent Browsing
Chapter 2: Troubleshooting Procedures
2.34 Intermittent Browsing
1. The Master Alarm can only be browsed intermittently.
Yes
No
L J
Go to “Final Actions” on page 2-7.
2. Ensure that the Ethernet connections are properly connected on all of the
alarm devices (refer to procedure 4.25 on page 4-68). They are properly con-
nected.
Yes
No
L J
Connect the Ethernet to all of the Alarm devices. If this solves
the problem, go to “Final Actions” on page 2-7. Otherwise, go
to step 3.
3. Check the LEDs on the Master Alarm Modules and the Area Alarm Network
Communications Boards. See “Master Alarm Power-Up” on page 1-60. Also
see “Area Alarm Power-Up” on page 1-112. The LEDs that indicate whether
the alarms are communicating are flashing very rapidly.
Yes
No
L J
Call your facility’s Information Systems personnel to fix the
problem. If they fix the problem, go to “Final Actions” on page
2-7. Otherwise, call BeaconMedæs Technical Support.
4. Have your facility’s Information Systems personnel check the
ARP
traffic of
the facility’s network. Heavy broadcast traffic can interfere with timely Web
service and data exchange between the Master Alarms and Area Alarm
Network Communications Boards. The broadcast traffic is heavy.
Yes
No
L J
Call BeaconMedæs Technical Support, then go to “Final
Actions” on page 2-7.
5. If there is excessive broadcast traffic, it usually means that there is a piece of
equipment on the facility’s network that is not operating properly. The facili-
ty’s Information Systems personnel will have to eliminate some of the traffic
flow, by fixing their other equipment. The browsing works correctly after the
Information Systems personnel fix their other equipment.
Yes
No
L J
Call BeaconMedæs Technical Support.
Summary of Contents for MedPlus TotalAlert
Page 370: ...Schematic Wiring Diagram Typical Local Alarm...
Page 371: ...Schematic Wiring Diagram Typical Master Alarm...
Page 372: ...Schematic Wiring Diagram Typical Master Alarm to Master Alarm...
Page 373: ...Schematic Wiring Diagram Typical Area Alarm...
Page 374: ...Schematic Wiring Diagram Typical Master Alarm to Three Area Alarm...
Page 375: ...Schematic Wiring Diagram Typical Master Alarm to Six Area Alarm Combination...