Page 2-42
MedPlus TotalAlert® Alarm Network Operation and Maintenance Manual (205728)
While Browsing the Master Alarm, Area Alarm Modules Are Not Accessible
Chapter 2: Troubleshooting Procedures
2.36 While Browsing the Master Alarm, Area Alarm Modules Are
Not Accessible
1. While you are browsing the Master Alarm, you cannot use the jump link to
go to Area Alarm Modules.
Yes
No
L J
Go to “Final Actions” on page 2-7.
2. Type in the word
Area
followed by
its
serial number in your Internet brows-
er’s address window, such as
Area26453
. The Area Alarm cannot be browsed.
Yes
No
L J
If this solved the problem, go to “Final Actions” on page 2-7.
Otherwise, go to step 3.
3. The Ethernet is connected properly.
Yes
No
L J
Ensure that the Ethernet is connected properly on the Area
Alarm Network Communications Board (refer to procedure 4.25
on page 4-68). If this solves the problem, go to “Final Actions”
on page 2-7. Otherwise, go to step 4.
4. Check the LEDs on the Area Alarm Network Communications Board. The
LEDs that indicate whether the alarms are communicating are flashing at a
steady rate.
Yes
No
L J
Call your facility’s Information Systems personnel to fix the
problem. If they fix the problem, go to “Final Actions” on page
2-7. Otherwise, go to step 5.
5. If the problem still exists, call BeaconMedæs Technical Support.
Summary of Contents for MedPlus TotalAlert
Page 370: ...Schematic Wiring Diagram Typical Local Alarm...
Page 371: ...Schematic Wiring Diagram Typical Master Alarm...
Page 372: ...Schematic Wiring Diagram Typical Master Alarm to Master Alarm...
Page 373: ...Schematic Wiring Diagram Typical Area Alarm...
Page 374: ...Schematic Wiring Diagram Typical Master Alarm to Three Area Alarm...
Page 375: ...Schematic Wiring Diagram Typical Master Alarm to Six Area Alarm Combination...