• Alarm or error logs into the Master Alarm’s memory.
• Electronic message is sent.
• If the Network Communications Board is installed in the Area Alarm, and the
Area Alarm Modules fail to respond to the Master Alarm; the Master Alarm
sounds, message is sent to the event log and displayed, and an electronic
message is sent.
During an alarm condition or a system error on the Area Alarm, the following
occurs:
• Area Alarm sounds an alarm, and the corresponding red high or low alarm
LED lights, if applicable (errors do not light LEDs).
• If the gas pressure is above or below its setpoints or if the vacuum is below
its setpoint, the pressure reading flashes.
• If it is an alarm error, the error code displays on the corresponding Area
Alarm Module.
• If the Network Communications Board is installed in the Area Alarm:
•
Area Alarm message is sent to the Master Alarm.
•
Master Alarm sounds.
•
Message displays on the Master Alarm.
•
Error or alarm logs into the event logs of the Master Alarm and the Area
Alarm.
•
Electronic message is sent from the Master Alarm.
All alarms include the following:
• Alarm silence key to silence the audible alarm.
• Red LEDs that remain lit until the alarm condition is no longer present.
• Test key to perform a self-test of each alarm.
MedPlus TotalAlert® Alarm Network Operation and Maintenance Manual (205728)
Page 3-5
Theory of Operation
Chapter 3: Theory of Operation
3
Summary of Contents for MedPlus TotalAlert
Page 370: ...Schematic Wiring Diagram Typical Local Alarm...
Page 371: ...Schematic Wiring Diagram Typical Master Alarm...
Page 372: ...Schematic Wiring Diagram Typical Master Alarm to Master Alarm...
Page 373: ...Schematic Wiring Diagram Typical Area Alarm...
Page 374: ...Schematic Wiring Diagram Typical Master Alarm to Three Area Alarm...
Page 375: ...Schematic Wiring Diagram Typical Master Alarm to Six Area Alarm Combination...