Page 4-10
MedPlus TotalAlert® Alarm Network Operation and Maintenance Manual (205728)
Master Alarm Module
Chapter 4: Removal, Replacement, and Adjustment Procedures
4.2
Master Alarm Module
Tools required:
Screwdriver
5/32" hex wrench
Removal
1. Using a 5/32" hex wrench, unlock the front panel (A) (see figure 4-3 on page
4–10).
Figure 4-3. Master Alarm Module
Shock Hazard:
Turn off the power supply in the alarm, and leave it off during the installa-
tion or replacement of any component or sub-assembly. Failure to do so
could result in personal injury or equipment damage.
2. Turn the power switch (F) to the OFF position.
3. Push open the ears (D) of the connectors to remove the ribbon cables (B).
4. Using a screwdriver, remove the four screws (C) that secure the Master
Alarm Module (E) to the front panel (A).