Page 4-12
MedPlus TotalAlert® Alarm Network Operation and Maintenance Manual (205728)
Master Alarm Battery
Chapter 4: Removal, Replacement, and Adjustment Procedures
4.3
Master Alarm Battery
Tools required:
5/32" hex wrench
Screwdriver
Removal
1. Using a 5/32" hex wrench, unlock the front panel (A) (see figure 4-4 on page
4–12).
Figure 4-4. Master Alarm Battery
Shock Hazard:
Turn off the power supply in the alarm, and leave it off during the installa-
tion or replacement of any component or sub-assembly. Failure to do so
could result in personal injury or equipment damage.
2. Turn the power switch (D) to the OFF position.
3. Slide a screwdriver underneath the battery (C), and lift it out of its holder (B).