Page 4-14
MedPlus TotalAlert® Alarm Network Operation and Maintenance Manual (205728)
Master Alarm Breakout P.C. Board
Chapter 4: Removal, Replacement, and Adjustment Procedures
4.4
Master Alarm Breakout P.C. Board
Tools required:
5/32" hex wrench
Needle nose pliers
Screwdriver
Removal
1. Using a 5/32" hex wrench, unlock the front panel (A) (see figure 4-5 on page
4–14).
Figure 4-5. Master Alarm Breakout P.C. Board
Shock Hazard:
Turn off the power supply in the alarm, and leave it off during the installa-
tion or replacement of any component or sub-assembly. Failure to do so
could result in personal injury or equipment damage.
2. Turn the power switch (G) to the OFF position.
3. Note the positions of the wiring at the connectors (I) for the source equip-
ment.
4. Using a screwdriver, loosen the screws in the connectors (I).