3. Push open the ears of the connectors (F) to remove the ribbon cables (B) and
(C).
4. If there is an Ethernet cable connected, unplug it.
5. Pull the P.C. board off of the mounting pegs (G).
Replacement
1. To replace the P.C. board, reverse the order of the removal instructions.
2. To ensure proper operation, perform the “Function Checks” on page 2–5.
MedPlus TotalAlert® Alarm Network Operation and Maintenance Manual (205728)
Page 4-25
Area Alarm Connector P.C. Board or Network Communications Board
Chapter 4: Removal, Replacement, and Adjustment Procedures
4