Figure 4-17. Power Supply Fuse Holder
3. Using a phillips head screwdriver, remove the three screws (F) that secure the
cover plate (G) to the back box (H).
4. Using needle nose pliers, remove the flag terminals (E) from the power sup-
ply fuse holder (D).
5. Using needle nose pliers, push down the ears on the power supply fuse holder
(D).
6. Slide the power supply fuse holder (D) out of the cover plate (G).
7. Using a screwdriver, remove the cover (B) of the power supply fuse holder
(D) and the fuse (C), and save them for the replacement procedure.
Replacement
1. To replace the power supply fuse holder, reverse the order of the removal
instructions.
2. To ensure proper operation, perform the “Function Checks” on page 2–5.
MedPlus TotalAlert® Alarm Network Operation and Maintenance Manual (205728)
Page 4-43
Power Supply Fuse Holder
Chapter 4: Removal, Replacement, and Adjustment Procedures