Page 4-46
MedPlus TotalAlert® Alarm Network Operation and Maintenance Manual (205728)
Adjustable Grip Latch
Chapter 4: Removal, Replacement, and Adjustment Procedures
4.18 Adjustable Grip Latch
Tools required:
5/32" hex wrench
11/16" wrench
Removal
1. Using a 5/32" hex wrench, unlock the lock (A) (see figure 4-19 on page
4–46).
Figure 4-19. Adjustable Grip Latch
Shock Hazard:
Turn off the power supply in the alarm, and leave it off during the installa-
tion of any component or sub-assembly. Failure to do so could result in
personal injury or equipment damage.
2. Turn the power switch (G) to the OFF position.
3. Grasp and hold the latch (B).
4. Using a 5/32" hex wrench, turn the lock (A) counterclockwise to back the
latch (B) down.