b. Remove and save the two screws (F) securing the power supply assembly
(C) to the mounting bracket (G).
c. Without putting excessive strain on the input power wiring (H) or the
ground wire (I), set the power supply assembly (C) to the side.
d. Remove and save the two screws (J) securing the mounting bracket (G) to
the rough-in box (K).
e. Remove and save the mounting bracket (G).
Figure 4-23. Transducer and DISS Demand Flow Valve
5. Note the position of the green wire (or shield wire) (L) in the terminal block
(M) located inside the transducer (A).
NOTE:
Polarity does not need to be observed for the red and black wires.
6. Loosen the screws (N) securing the three wires (L and O) to the terminal
block (M).
7. Remove the wires (L and O) from the terminal block (M).
8. Disconnect the DISS nut (P) from the DISS demand flow valve (B).
9. Remove the transducer (A).
MedPlus TotalAlert® Alarm Network Operation and Maintenance Manual (205728)
Page 4-59
Local-Mount Transducer and DISS Demand Flow Valve Assembly for the Conversion Kit Housed in the
Area Line Pressure Alarm (0321) Rough-In Box
Chapter 4: Removal, Replacement, and Adjustment Procedures
4