Preventive Maintenance Schedule
Permanent records of all test results for alarm systems should be made at the
time of installation and at all required intervals as specified by NFPA 99 (see
table 6-1 on page 6–31).
Table 6-1. Preventive Maintenance Schedule
Contact BeaconMedæs Technical Support for information on having these tests performed by a
BeaconMedæs Service Representative.
Function
Procedure
Gas pressures
Check the readings on the area alarm module digital dis-
plays daily.
Audible alarm signals
Test once a month by pressing the test keys on each
front panel (see “Master Alarm Self-Test” on page 2-5),
(see “Area Alarm Self-Test” on page 2-6), and (see
“Local Alarm Self-Test” on page 2-6).
Visual alarm signals
Test once a month by pressing the test keys on each
front panel (see “Master Alarm Self-Test” on page 2-5),
(see “Area Alarm Self-Test” on page 2-6), and (see
“Local Alarm Self-Test” on page 2-6).
Transducer visual indicators
Inspect once a month (see “Transducer” on page 2-11).
Alarm system components
Check every alarm component for wear.
Power and signal wiring
Inspect all wiring for good condition.
Electrical connections
Check all electrical connections for tightness, and tight-
en as needed.
Gas hosesw (on local-mount
transducers)
Inspect hoses for cracking, wear, collapse, or leakage
annually.
General appearance
Check cleanliness. Clean if necessary, see “Cleaning” on
page 6-28.
MedPlus TotalAlert® Alarm Network Operation and Maintenance Manual (205728)
Page 6-31
Preventive Maintenance
Chapter 6: General Procedures
6
Summary of Contents for MedPlus TotalAlert
Page 370: ...Schematic Wiring Diagram Typical Local Alarm...
Page 371: ...Schematic Wiring Diagram Typical Master Alarm...
Page 372: ...Schematic Wiring Diagram Typical Master Alarm to Master Alarm...
Page 373: ...Schematic Wiring Diagram Typical Area Alarm...
Page 374: ...Schematic Wiring Diagram Typical Master Alarm to Three Area Alarm...
Page 375: ...Schematic Wiring Diagram Typical Master Alarm to Six Area Alarm Combination...