Page 1-32
MedPlus TotalAlert® Alarm Network Operation and Maintenance Manual (205728)
Detailed Component Descriptions
Chapter 1: Introduction
There is one dry-contact relay featured in the Area Alarm that provides an output
signal. The output can be wired to either the normally-closed or normally-open
contacts. It is located on the Area Alarm Breakout P.C. Board, and marked MAIN
ALARM. For wiring diagrams, refer to the foldouts at the rear of this manual.
• MAIN ALARM RELAY
•
When the Master Alarm is in alarm condition, this relay activates.
•
When the Master Alarm is silenced, this relay stays activated.
•
When the Master Alarm condition is remedied, this relay deactivates.
When the battery needs changed,
Real Time clock battery low
displays on the
Area Alarm’s
Home
page. The buzzer does not sound. This message logs into the
Error Log when the CheckSum Test is run, which is a test that the Area Alarm
runs automatically every 24 hours.
The
Silence key
silences the audible alarm. You can set the time between the
audible
alarms to resound never, every 30 minutes, every 60 minutes, every 90
minutes, or every 120 minutes. The alarm code remains on the display, the red
LED remains steadily lit, and the reading continues to flash indicating that the
alarm condition has not been remedied.
NOTE:
To comply with the requirements of the City of New York Department of
Buildings; Materials and Equipment Acceptance Division (M.E.A.), do not set
the alarm return time to “never.” Set the alarm to return every 30 minutes.
A
Test key
starts a self-diagnostic routine.
• The seven segment displays light each individual segment.
• The LEDs light.
• The audible alarm sounds.
Summary of Contents for MedPlus TotalAlert
Page 370: ...Schematic Wiring Diagram Typical Local Alarm...
Page 371: ...Schematic Wiring Diagram Typical Master Alarm...
Page 372: ...Schematic Wiring Diagram Typical Master Alarm to Master Alarm...
Page 373: ...Schematic Wiring Diagram Typical Area Alarm...
Page 374: ...Schematic Wiring Diagram Typical Master Alarm to Three Area Alarm...
Page 375: ...Schematic Wiring Diagram Typical Master Alarm to Six Area Alarm Combination...