Specifications
For MedPlus TotalAlert® Alarm Network specifications, refer to table 1-10 on
page 1-41 through table 1-15 on page 1-42.
Refer to table 1-16 on page 1-43 for the types of existing alarms that can be con-
verted to a MedPlus TotalAlert® Alarm Network—Master Alarm, MedPlus
TotalAlert® Alarm Network—Area Alarm, or MedPlus TotalAlert® Alarm
Network—Combination Alarm. Also included in the table are the sizes that the
existing rough-in boxes need to be, and the conversion kits that replace those
existing alarms.
Table 1-10. Power Specifications
Table 1-11. Cabling Requirements from the Area Alarm Module to the
Transducer
Specifications
Dimensions
Distance
Maximum 4500' (1372 m)
Cable
Stranded, 22-gauge, shielded, twisted-
pair, three-conductor cable
Signal Power
15 V DC, 75 mA maximum
Facility Power Specifications
Ratings
Supply voltage for single power supply alarms 120 V AC, 60 Hz
Breaker for single power supply alarms
15 A
Supply voltage for dual power supply alarms
120 V AC, 60 Hz
Breaker for dual power supply alarms
15 A
Field power supply wiring
12 gauge
Fuse rating for internal power supply (except
for the Master Alarm plus Six Area Alarm
Modules)
0.5 Amp Slo-Blo®a
Fuse rating for internal power supply on the
Master Alarm plus Six Area Alarm Modules
0.75 Amp Slo-Blo®
MedPlus TotalAlert® Alarm Network Operation and Maintenance Manual (205728)
Page 1-41
Specifications
Chapter 1: Introduction
1
Summary of Contents for MedPlus TotalAlert
Page 370: ...Schematic Wiring Diagram Typical Local Alarm...
Page 371: ...Schematic Wiring Diagram Typical Master Alarm...
Page 372: ...Schematic Wiring Diagram Typical Master Alarm to Master Alarm...
Page 373: ...Schematic Wiring Diagram Typical Area Alarm...
Page 374: ...Schematic Wiring Diagram Typical Master Alarm to Three Area Alarm...
Page 375: ...Schematic Wiring Diagram Typical Master Alarm to Six Area Alarm Combination...