The label strips read from left to right with “IN” on the left-hand side, and
“OUT” on the right-hand side.
6. Starting with the word “IN,” slide the label strip (BY) into the label opening
(BZ) with the words facing away from the rough-in assembly (CC).
7. Turn the power switch (CB) to the ON position.
8. Close the front panel (CA).
9. Using a 5/32" hex wrench, lock the front panel (CA).
10. For more information about installing an original alarm network, refer to the
MedPlus TotalAlert® Alarm Network Installation Instructions
.
11. For more information about installing a conversion kit alarm network, refer to
the
MedPlus TotalAlert® Alarm Network Conversion Kits Installation
Instructions
.
MedPlus TotalAlert® Alarm Network Operation and Maintenance Manual (205728)
Page 1-59